The SampSubtract component does not work correctly
Tech Note: TN0856
Product: RPvdsEx
Version: > 5.2
Date Added: 2009-06-02
Issue
When running a circuit containing the SampSubtract component on any single-DSP device (RP2.1, RA16, etc.) the SampSubtract component does not work correctly (the buffer is all zeros and never changes).
Workaround
The SampSubtract component is designed for multi-DSP devices such as the RX5 and RZ2, for single-DSP devices, contact tech support for a possible alternative solution.